Configuration
R&S
®
FSW-K7
126User Manual 1173.9240.02 ─ 17
on page 124) and "Phones Output", adjust the volume setting using the rotary knob on
the front panel.
Remote command:
OUTPut:ADEMod[:ONLine]:PHONes on page 231
5.10 Automatic Settings
Some settings can be adjusted by the R&S FSW automatically according to the current
measurement settings.
To activate the automatic adjustment of a setting, select the corresponding function in
the AUTO SET menu or in the configuration dialog box for the setting, where available.
MSRA/MSRT operating mode
In MSRA/MSRT operating mode, settings related to data acquisition cannot be adjus-
ted for Analog Demodulation applications.
Adjusting all Determinable Settings Automatically (Auto All)...................................... 126
Adjusting the Center Frequency Automatically (Auto Freq)........................................ 126
Setting the Reference Level Automatically (Auto Level).............................................127
Resetting the Automatic Measurement Time (Meastime Auto)...................................127
Changing the Automatic Measurement Time (Meastime Manual).............................. 127
Upper Level Hysteresis...............................................................................................127
Lower Level Hysteresis...............................................................................................127
AF Auto Scale............................................................................................................. 128
Adjusting all Determinable Settings Automatically (Auto All)
Activates all automatic adjustment functions for the current measurement settings.
This includes:
Auto Frequency
Auto Level
"AF Auto Scale" on page 119
This function is only available for the MSRA/MSRT Master, not for the applications.
Remote command:
[SENSe:]ADJust:ALL on page 267
Adjusting the Center Frequency Automatically (Auto Freq)
This function adjusts the center frequency automatically.
The optimum center frequency is the frequency with the highest S/N ratio in the fre-
quency span. As this function uses the signal counter, it is intended for use with sinus-
oidal signals.
This function is not available for input from the Digital Baseband Interface (R&S FSW-
B17).
Remote command:
[SENSe:]ADJust:FREQuency on page 268
Automatic Settings