Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
267User Manual 1173.9240.02 ─ 17
11.4.9 Adjusting Settings Automatically
The following remote commands are required to adjust settings automatically in a
remote environment. The tasks for manual operation are described in chapter 5.10,
"Automatic Settings", on page 126.
MSRA/MSRT operating mode
In MSRA/MSRT operating mode, settings related to data acquisition cannot be adjus-
ted for Analog Demodulation applications.
[SENSe:]ADJust:ALL......................................................................................................267
[SENSe:]ADJust:CONFigure:DURation............................................................................ 267
[SENSe:]ADJust:CONFigure:DURation:MODE..................................................................268
[SENSe:]ADJust:FREQuency..........................................................................................268
[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer..............................................................268
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer...............................................................269
[SENSe:]ADJust:LEVel...................................................................................................269
[SENSe:]ADJust:SCALe:Y:AUTO[:CONTinuous]...............................................................269
[SENSe:]ADJust:ALL
This command initiates a measurement to determine and set the ideal settings for the
current task automatically (only once for the current measurement).
This includes:
●
Center frequency
●
Reference level
●
Scaling
Example:
ADJ:ALL
Usage: Event
Manual operation: See "Adjusting all Determinable Settings Automatically (Auto
All)" on page 126
[SENSe:]ADJust:CONFigure:DURation <Duration>
In order to determine the ideal reference level, the R&S FSW performs a measurement
on the current input data. This command defines the length of the measurement if
[SENSe:]ADJust:CONFigure:DURation:MODE is set to MANual.
Parameters:
<Duration> Numeric value in seconds
Range: 0.001 to 16000.0
*RST: 0.001
Default unit: s
Configuring the Measurement