Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
266User Manual 1173.9240.02 ─ 17
Parameters:
<Value> *RST: AM time domain: 0 PCT; FM time domain: 0 Hz;
PM time domain: 0 rad; AM spectrum: 100 PCT;
FM spectrum: 250 kHz; PM spectrum: 10 rad;
Example:
DISP:TRAC:Y:RVAL 0
Sets the value assigned to the reference position to 0 Hz
Manual operation: See "Reference Value" on page 118
11.4.8.7 Scaling for RF Evaluation
These commands are required for RF evaluations and the result summary.
● DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:RPOSition on page 239
● DISPlay[:WINDow<n>]:TRACe:Y:SPACing on page 239
● DISPlay[:WINDow<n>]:TRACe:Y[:SCALe] on page 238
● DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:MODE on page 238
11.4.8.8 Units
The units define how the demodulated data is displayed.
UNIT:ANGLe................................................................................................................. 266
UNIT:THD..................................................................................................................... 266
UNIT:ANGLe <Unit>
This command selects the unit for angles (for PM display).
This command is identical to CALC:UNIT:ANGL
Parameters:
<Unit> DEG | RAD
*RST: RAD
Example:
UNIT:ANGL DEG
Manual operation: See "Phase Unit (Rad/Deg)" on page 121
UNIT:THD <Mode>
Selects the unit for THD measurements.
This command is identical to CALC:UNIT:THD
Parameters:
<Mode> DB | PCT
*RST: DB
Example:
UNIT:THD PCT
Manual operation: See "THD Unit (% / DB)" on page 121
Configuring the Measurement