Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
299User Manual 1173.9240.02 ─ 17
File contents Description
x-Axis;LIN; Scaling of x-axis linear (LIN) or logarithmic (LOG)
y-Axis;LOG; Scaling of y-axis linear (LIN) or logarithmic (LOG)
x-Unit;Hz; Unit of x values: Hz with span > 0; s with span = 0; dBm/dB
with statistics measurements
y-Unit;dBm; Unit of y values: dB*/V/A/W depending on the selected unit
with y-axis LOG or % with y-axis LIN
Data section for individual window
Window;1;Frequency Sweep Window number and name
Trace 1;; Selected trace
Trace Mode;AVERAGE; Display mode of trace: CLR/WRITE,AVER-
AGE,MAXHOLD,MINHOLD
Detector;AUTOPEAK; Detector set: AUTOPEAK,MAXPEAK,MINPEAK,AVER-
AGE,RMS,SAMPLE,QUASIPEAK
Values; 1001; Number of measurement points
10000;-10.3;-15.7
10130;-11.5;-16.9
10360;-12.0;-17.4
...;...;
Measured values: <x value>, <y1>, <y2>; <y2> being available
only with detector AUTOPEAK and containing in this case the
smallest of the two measured values for a measurement point.
Data section for individual trace
Trace 2;; Next trace in same window
...
Data section for individual window
Window;2 ..; Name of next window
Data section for individual trace
Trace 1;; First trace
...
11.8 Analyzing Results
The following remote commands are required to configure general result analysis set-
tings concerning the trace, markers, lines etc. in a remote environment. They are iden-
tical to the analysis functions in the base unit except for some special marker functions
and spectrograms, which are not available in the Analog Demodulation application.
More details are described for manual operation in chapter 6, "Analysis", on page 129.