Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
176User Manual 1173.9240.02 ─ 17
Suffix Value range Description
<n> 1..6 Window or Evaluation
<t> 1..6 Trace
11.3 Activating Analog Demodulation Measurements
Analog demodulation measurements require a special application on the R&S FSW.
The measurement is started immediately with the default settings.
INSTrument:CREate:DUPLicate...................................................................................... 176
INSTrument:CREate[:NEW]............................................................................................ 176
INSTrument:CREate:REPLace........................................................................................177
INSTrument:DELete....................................................................................................... 177
INSTrument:LIST?......................................................................................................... 177
INSTrument:REName.....................................................................................................179
INSTrument[:SELect]......................................................................................................179
SYSTem:PRESet:CHANnel[:EXECute]............................................................................ 179
INSTrument:CREate:DUPLicate
This command duplicates the currently selected measurement channel, i.e starts a
new measurement channel of the same type and with the identical measurement set-
tings. The name of the new channel is the same as the copied channel, extended by a
consecutive number (e.g. "Spectrum" -> "Spectrum 2").
The channel to be duplicated must be selected first using the INST:SEL command.
This command is not available if the MSRA/MSRT Master channel is selected.
Example:
INST:SEL 'Spectrum'
INST:CRE:DUPL
Duplicates the channel named 'Spectrum' and creates a new
measurement channel named 'Spectrum 2'.
Usage: Event
INSTrument:CREate[:NEW] <ChannelType>, <ChannelName>
This command adds an additional measurement channel. The number of measure-
ment channels you can configure at the same time depends on available memory.
Parameters:
<ChannelType> Channel type of the new channel.
For a list of available channel types see INSTrument:LIST?
on page 177.