Measurement Example: Demodulating an FM Signal
R&S
®
FSW-K7
163User Manual 1173.9240.02 ─ 17
9 Measurement Example: Demodulating an
FM Signal
A practical example for a basic Analog Demodulation measurement is provided here. It
demonstrates how operating and measurement errors can be avoided using correct
configuration settings.
The measurement is performed using the following devices:
An R&S FSW with application firmware R&S FSW-K7: Analog Demodulation
A vector signal generator, e.g. R&S SMU
Signal
Generator
R&S FSW
Fig. 9-1: Test setup
Signal generator settings (e.g. R&S SMU):
Frequency: 500 MHz
Level: -10 dBm
Modulation: FM
Modulation frequency: 10 kHz
Frequency deviation: 50 kHz
Procedure:
1.
Preset the R&S FSW.
2.
Set the center frequency to 500
MHz.
3.
Set the reference level to 0
dBm.
4. Select the MODE key and then the "Analog Demod" button.
By default, the FM Time Domain result display and a Result Summary are shown.