Analysis
R&S
®
FSW-K7
132User Manual 1173.9240.02 ─ 17
"Linear"
The power level values are converted into linear units prior to averag-
ing. After the averaging, the data is converted back into its original
unit.
"Logarithmic"
For logarithmic scaling, the values are averaged in dBm. For linear
scaling, the behavior is the same as with linear averaging.
"Power"
Activates linear power averaging.
The power level values are converted into unit Watt prior to averag-
ing. After the averaging, the data is converted back into its original
unit.
Use this mode to average power values in Volts or Amperes cor-
rectly.
Remote command:
[SENSe:]AVERage<n>:TYPE on page 275
Average Count
Determines the number of averaging or maximum search procedures If the trace
modes "Average", "Max Hold" or "Min Hold" are set.
In continuous sweep mode, if sweep count = 0 (default), averaging is performed over
10 sweeps. For sweep count =1, no averaging, maxhold or minhold operations are per-
formed.
Remote command:
[SENSe:]AVERage:COUNt on page 274
Predefined Trace Settings - Quick Config
Commonly required trace settings have been predefined and can be applied very
quickly by selecting the appropriate button.
Function Trace Settings
Preset All Traces Trace 1: Clear Write
Traces 2-6: Blank
Set Trace Mode
Max | Avg | Min
Trace 1: Max Hold
Trace 2: Average
Trace 3: Min Hold
Traces 4-6: Blank
Set Trace Mode
Max | ClrWrite | Min
Trace 1: Max Hold
Trace 2: Clear Write
Trace 3: Min Hold
Traces 4-6: Blank
Trace 1/Trace 2/Trace 3/Trace 4 (Softkeys)
Displays the "Traces" settings and focuses the "Mode" list for the selected trace.
Remote command:
DISPlay[:WINDow<n>]:TRACe<t>[:STATe] on page 272
Trace Settings