Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
272User Manual 1173.9240.02 ─ 17
Parameters:
<State> ON
The automatic reset is off.
OFF
The automatic reset is on.
*RST: OFF
Example:
DISP:WIND:TRAC3:MODE:HCON ON
Switches off the reset function.
Manual operation: See "Hold" on page 131
DISPlay[:
WINDow<n>]:TRACe<t>:SELect
This command selects the trace specified by the index <t> in the window specified by
the index <n>. Only traces that are active in the specified result display can be
selected. The selected trace is used to determine the "Result Summary" for the corre-
sponding result display (see "Result Summary" on page 23).
The query returns the number of the currently selected trace in the window specified by
the index <n> (trace index is ignored). Traces can only be queried for graphical result
displays (not Result Summary, Marker Table or Peak Marker List).
Return values:
<TraceNo> Number of the currently selected trace.
Example:
DISP:TRAC3:SEL
Usage: SCPI confirmed
DISPlay[:WINDow<n>]:TRACe<t>[:STATe] <State>
This command turns a trace on and off.
The measurement continues in the background.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1 for TRACe1, 0 for TRACe 2 to 6
Example:
DISP:TRAC3 ON
Usage: SCPI confirmed
Manual operation: See "Trace 1/Trace 2/Trace 3/Trace 4 (Softkeys)" on page 132
[SENSe:]ADEMod:AM[:ABSolute][:TDOMain][:TYPE]
[SENSe:]ADEMod:AM:RELative[:TDOMain][:TYPE]
[SENSe:]ADEMod:AM:RELative:AFSPectrum[:TYPE]
[SENSe:]ADEMod:FM[:TDOMain][:TYPE]
[SENSe:]ADEMod:FM:AFSPectrum[:TYPE]
[SENSe:]ADEMod:PM[:TDOMain][:TYPE]
Configuring the Measurement