Configuration
R&S
®
FSW-K7
56User Manual 1173.9240.02 ─ 17
Some settings are also available in the "Amplitude" tab of the "Amplitude" dialog box.
Since the Digital I/Q input and the Analog Baseband input use the same digital signal
path, both cannot be used simultaneously. When one is activated, established connec-
tions for the other are disconnected. When the second input is deactivated, connec-
tions to the first are re-established. This may cause a short delay in data transfer after
switching the input source.
External mixers are not supported in MSRA/MSRT mode.
Radio Frequency Input............................................................................................56
External Mixer Settings........................................................................................... 58
Digital I/Q Input Settings......................................................................................... 69
Analog Baseband Input Settings.............................................................................71
Probe Settings.........................................................................................................72
External Generator Control Settings....................................................................... 73
5.4.1.1 Radio Frequency Input
The default input source for the R&S FSW is "Radio Frequency", i.e. the signal at the
RF INPUT connector on the front panel of the R&S FSW. If no additional options are
installed, this is the only available input source.
Radio Frequency State................................................................................................. 56
Input Coupling...............................................................................................................56
Impedance.................................................................................................................... 57
High-Pass Filter 1...3 GHz............................................................................................ 57
YIG-Preselector.............................................................................................................57
Input Connector.............................................................................................................58
Radio Frequency State
Activates input from the RF INPUT connector.
Remote command:
INPut:SELect on page 184
Input Coupling
The RF input of the R&S FSW can be coupled by alternating current (AC) or direct cur-
rent (DC).