Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
275User Manual 1173.9240.02 ─ 17
Usage: SCPI confirmed
Manual operation: See "Sweep/Average Count" on page 106
See "Average Count" on page 132
[SENSe:]AVERage<n>[:STATe<t>] <State>
This command turns averaging for a particular trace in a particular window on and off.
Parameters:
<State> ON | OFF
Usage: SCPI confirmed
[SENSe:]AVERage<n>:TYPE <Mode>
This command selects the trace averaging mode.
Parameters:
<Mode> VIDeo
The logarithmic power values are averaged.
LINear
The power values are averaged before they are converted to
logarithmic values.
POWer
The power level values are converted into unit Watt prior to
averaging. After the averaging, the data is converted back into
its original unit.
*RST: VIDeo
Example:
AVER:TYPE LIN
Switches to linear average calculation.
Usage: SCPI confirmed
Manual operation: See "Average Mode" on page 131
[SENSe:][WINDow<n>:]DETector<trace>[:FUNCtion] <Detector>
Defines the trace detector to be used for trace analysis.
Configuring the Measurement