Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
276User Manual 1173.9240.02 ─ 17
Parameters:
<Detector> APEak
Autopeak
NEGative
Negative peak
POSitive
Positive peak
QPEak
Quasipeak (CISPR filter only)
SAMPle
First value detected per trace point
RMS
RMS value
AVERage
Average
CAVerage
CISPR Average (CISPR filter only)
CRMS
CISPR RMS (CISPR filter only)
*RST:
APEak (I/Q Analyzer: RMS)
Example:
DET POS
Sets the detector to "positive peak".
Manual operation: See "Detector" on page 131
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]:AUTO <State>
This command couples and decouples the detector to the trace mode.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
DET:AUTO OFF
The selection of the detector is not coupled to the trace mode.
Manual operation: See "Detector" on page 131