Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
238User Manual 1173.9240.02 ─ 17
Parameters:
<Gain> 15 dB | 30 dB
The availability of preamplification levels depends on the
R&S FSW model.
R&S FSW8/13: 15dB and 30 dB
R&S FSW26 or higher: 30 dB
All other values are rounded to the nearest of these two.
*RST: OFF
Example:
INP:GAIN:VAL 30
Switches on 30 dB preamplification.
Usage: SCPI confirmed
Manual operation: See "Preamplifier (option B24)" on page 89
11.4.5.4 Scaling the Y-Axis
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]........................................................................238
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:AUTO ONCE....................................................238
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:MODE............................................................. 238
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:RPOSition........................................................239
DISPlay[:WINDow<n>]:TRACe:Y:SPACing.......................................................................239
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe] <Range>
This command defines the display range of the y-axis.
Parameters:
<Range> If the y-axis shows the power, the unit is dB with a range from
10 dB to 200 dB.
If the y-axis shows the frequency, the unit is Hz with a variable
range.
*RST: 100 dB (frequency domain), 500 kHz (time domain)
Example:
DISP:TRAC:Y 110dB
Usage: SCPI confirmed
Manual operation: See "Range" on page 120
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:AUTO ONCE
Automatic scaling of the y-axis is performed once, then switched off again.
Usage:
SCPI confirmed
Manual operation: See "Auto Scale Once" on page 120
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:MODE <Mode>
This command selects the type of scaling of the y-axis.
Configuring the Measurement