EasyManua.ls Logo

Rohde & Schwarz FSW Series - [SENSe:]ADEMod:AM[:ABSolute][:TDOMain]:RESult?

Rohde & Schwarz FSW Series
371 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
290User Manual 1173.9240.02 ─ 17
[SENSe:]ADEMod:AM[:ABSolute][:TDOMain]:RESult? <TraceMode>
[SENSe:]ADEMod:AM:RELative[:TDOMain]:RESult? <TraceMode>
[SENSe:]ADEMod:AM:RELative:AFSPectrum:RESult? <TraceMode>
[SENSe:]ADEMod:FM[:TDOMain]:RESult? <TraceMode>
[SENSe:]ADEMod:FM:AFSPectrum:RESult? <TraceMode>
[SENSe:]ADEMod:PM[:TDOMain]:RESult? <TraceMode>
[SENSe:]ADEMod:PM:AFSPectrum:RESult? <TraceMode>
[SENSe:]ADEMod:SPECtrum:RESult? <TraceMode>
This command reads the result data of the evaluated signal in the specified trace
mode. The data format of the output data block is defined by the FORMat command
(see FORMat[:DATA] on page 291).
The trace results are configured for a specific evaluation. The following table indicates
which command syntax refers to which evaluation method, as well as the output unit of
the results.
Command syntax Evaluation method Output unit
AM[:ABSolute][:TDOMain] RF time domain dBm
AM:RELative[:TDOMain] AM time domain %
AM:RELative:AFSPectrum AM spectrum %
FM[:TDOMain] FM time domain kHz
FM:AFSPectrum FM spectrum kHz
PM[:TDOMain] PM time domain rad or °
PM:AFSPectrum PM spectrum rad or °
SPECtrum RF spectrum dBm (logarithmic display)
or V (linear display).
Query parameters:
<TraceMode> WRITe | AVERage | MAXHold | MINHold | VIEW
The specified trace mode must be one of those configured by
SENS:ADEM:<Evaluation>:TYPE, see [SENSe:]ADEMod:
SPECtrum[:TYPE] on page 273. Otherwise a query error is
generated.
Example:
ADEM:AM AVER,MAXH,MINH
Sets up RF time domain results to be measured
INIT; *WAI
Starts measurement and waits for sync
FORM ASC
Selects output format
ADEM:AM:RES? AVER
Reads RF time domain average results
ADEM:AM:RES? MAXH
Reads RF time domain max hold results
ADEM:AM:RES? MINH
Reads RF time domain min hold results
Retrieving Results