Remote Commands for the I/Q Analyzer
R&S
®
FSWP
136User Manual 1177.5856.02 ─ 07
160 MHz | MAX
The bandwidth extension option is activated. The currently avail-
able maximum bandwidth is allowed.
TRACe:IQ:WBANd[:STATe] is set to ON.
*RST: maximum available
Default unit: Hz
Example:
TRAC:IQ:WBAN:MBW 82 MHZ
TRAC:IQ:WBAN:MBW?
160000000
Manual operation: See " Maximum Bandwidth " on page 59
TRACe:IQ:WBANd[:STATe]
<State>
This command determines whether the wideband provided by bandwidth extension
options is used or not (if installed).
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Manual operation: See " Maximum Bandwidth " on page 59
TRACe:IQ:WFILter <State>
Activates a 200 MHz filter before the A/D converter, thus restricting the processed
bandwidth to 200 MHz while using the wideband processing path in the R&S FSWP.
For prerequisites see manual operation.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST: 0
Manual operation: See "200 MHz Filter" on page 60
8.6.12 Automatic Configuration
[SENSe:]ADJust:ALL......................................................................................................137
[SENSe:]ADJust:CONFigure:DURation............................................................................ 137
[SENSe:]ADJust:CONFigure:DURation:MODE..................................................................137
[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer.............................................................. 138
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer...............................................................138
Configuring I/Q Analyzer Measurements