Remote Commands for the I/Q Analyzer
R&S
®
FSWP
137User Manual 1177.5856.02 ─ 07
[SENSe:]ADJust:CONFigure:TRIGger..............................................................................138
[SENSe:]ADJust:FREQuency..........................................................................................139
[SENSe:]ADJust:LEVel................................................................................................... 139
[SENSe:]ADJust:ALL
This command initiates a measurement to determine and set the ideal settings for the
current task automatically (only once for the current measurement).
This includes:
●
Reference level
Example:
ADJ:ALL
Manual operation: See " Adjusting all Determinable Settings Automatically ( Auto
All )" on page 66
[SENSe:]ADJust:CONFigure:DURation <Duration>
In order to determine the ideal reference level, the R&S FSWP performs a measure-
ment on the current input data. This command defines the length of the measurement
if [SENSe:]ADJust:CONFigure:DURation:MODE is set to MANual.
Parameters:
<Duration> Numeric value in seconds
Range: 0.001 to 16000.0
*RST: 0.001
Default unit: s
Example:
ADJ:CONF:DUR:MODE MAN
Selects manual definition of the measurement length.
ADJ:CONF:LEV:DUR 5ms
Length of the measurement is 5 ms.
Manual operation: See " Changing the Automatic Measurement Time ( Meastime
Manual )" on page 67
[SENSe:]ADJust:CONFigure:DURation:MODE <Mode>
In order to determine the ideal reference level, the R&S FSWP performs a measure-
ment on the current input data. This command selects the way the R&S FSWP deter-
mines the length of the measurement .
Parameters:
<Mode> AUTO
The R&S FSWP determines the measurement length automati-
cally according to the current input data.
MANual
The R&S FSWP uses the measurement length defined by
[SENSe:]ADJust:CONFigure:DURation on page 137.
*RST: AUTO
Configuring I/Q Analyzer Measurements