109
Operating the Scale
Practical Example
Differential weighing: Combined weighing; create lot, determine the difference in weight between initial weights and
backweights of three samples (with autoprint of the formatted backweighing record)
Settings (changes in the factory settings required for this example):
Setup: App(lication parameters): Application 1: Differential weighing: Combined weighing
Step Press key(s) Display/Output
(or follow instructions)
1. Turn the scale on, if necessary e
2. Tare the scale, if necessary t
3. Start combined weighing
Start soft key
4. Select lot ID input Lot soft key
5. Enter lot ID a 1 ... 9 0 .
6. Confirm input l soft key
7. Activate weight readout oo soft key
(or toggle to combined
Wt.Seq soft key
weighing)
8. Measure 1st tare weight Place 1st empty container
on scale