EasyManuals Logo

Sartorius Factory Series User Manual

Sartorius Factory Series
213 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #153 background imageLoading...
Page #153 background image
153
Operating the Scale
Practical Example 3: Calculation with statistics
Statistically determine the average gsm weight of A4 paper and document the result with a printout of the results
on 10 samples. The gsm weight is a product of the division of the weight by the surface area. One A4 sheet has
a surface area of 0.210 m x 0.297 m = 0.06237 m
2
.
Settings (changes in the factory settings required for this example):
Setup: App(lication parameters): Application 1: Calculation: Decimal places in calculated result: 3 decimal places
Setup: App(lication parameters): Application 2: Off
Setup: App(lication parameters): Application 3: Statistics: Automatic storage: On, first value at stability
Setup: App(lication parameters): Application 3: Statistics: Evaluated value: Calculated
Setup: App(lication parameters): Application 3: Statistics: M+/M function, then tare: On
Step Key (or instruction) Display/Output
1. If necessary: turn on the scale e
and enter the settings given above
2. Clear Statistics memory and c
equation memory, if necessary
3. Place a container for the paper t
on the scale and tare
4. Toggle to Calculation D
5. Select equation input Equat. soft key
6. Enter equation
Weight soft key
(here: EQUAT.=W*0.06237) / soft key
0 . 0 6 2
3 7
7. Exit the equation input mode Start soft key
8. Toggle to Application 3: D
Statistics

Table of Contents

Other manuals for Sartorius Factory Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Factory Series and is the answer not in the manual?

Sartorius Factory Series Specifications

General IconGeneral
BrandSartorius
ModelFactory Series
CategoryScales
LanguageEnglish

Related product manuals