Operating Instructions Flexsafe
®
Pro Mixer Drive Unit 29
Installation
5.4 Removing the Transport Lock
The device is protected with a transport lock on delivery. The transport lock
consists of 2 transport lock screws. The transport lock screws must be
removed in order to operate the device.
Personnel: 2 persons, user
Procedure
t Hold the device firmly by the handle and push it down in a controlled
manner.
t Push the device down until the handle is resting on the floor.
t Have a second person hold the device firmly by the handle, so the device
remains in this position during installation
t Unscrew the transport lock screws (1) and (2). The transport lock screws
are marked in red.
t Put the transport lock screws aside should the device ever need to be
transported again later.
t Screw the 2 supplied screws into the device as a replacement for the
removed transport lock screws.
t While holding the device by the handle, tip it downwards once again in a
controlled manner so that it stands on the castors.
t
CAUTION Risk of injury due to uncontrolled rolling away of the
device! Place it on a level surface and secure the brakes. Wear personal
protective equipment.
2
1