Operating Instructions Flexsafe
®
Pro Mixer Drive Unit 53
Decommissioning
11 Decommissioning
11.1 Decommissioning the Device
Requirements
The device is removed from the Palletank
®
(see Chapter “8.12 Ending the
Mixing Process”, page 45).
Procedure
t Turn the device off.
t Disconnect the power supply cable from the power supply at the instal-
lation site.
t Disconnect the potential equalization cable from the power supply at
the installation site.
t Unscrew the nut for the strain relief (1).
t Remove the power supply cable from the cable holder.
t Screw in the nut for the strain relief.
t Remove the potential equalization cable from the “Potential
equalization” connection.
t If the device is connected to a remote control: Remove the connection
cable and seal the connection with the locking cover cap.
t
CAUTION Risk of injury due to uncontrolled rolling away of the
device! Place it on a level surface and secure the brakes. Wear personal
protective equipment.
1

Table of Contents