30 Operating Instructions Flexsafe
®
Pro Mixer Drive Unit
Getting Started
6 Getting Started
6.1 Connecting the Power Supply
6.1.1 Connecting a Potential Equalization Cable
Procedure
t
WARNING Risk of injury due to electric shock when using an unsuita-
ble potential equalization cable! Use the potential equalization cable
provided. The potential equalization cable is combined with the power
supply cable.
t Check whether the mains plug of the potential equalization cable
matches the power connections at the installation site.
t If necessary: Contact Sartorius Service.
t Connect the potential equalization cable to the “Potential equalization”
connection (1).
t Connect the mains plug of the potential equalization cable to the power
supply provided at the installation site.
6.1.2 Connecting the Power Supply Cable
Procedure
t
WARNING Risk of injury due to electric shock when using an unsuit-
able power supply cable! Use the power supply cable provided.
t Check whether the country-specific power plug matches the power con-
nections at the installation site.
t If required: Contact Sartorius Service (for available power supply
cables, see Chapter “16 Accessories”, page 60).
t Unscrew the nut for the strain relief (2) and remove.
t Insert the power supply cable into the cable holder.
t Reinsert the nut for the strain relief and tighten by hand.
t Press the grooved surface on the plug of the power supply cable and
connect the plug to the “power supply” connection (1).
t Connect the mains plug of the power supply cable to the power supply
provided at the installation site.