46 Operating Instructions Flexsafe
®
Pro Mixer Drive Unit
Operation
Procedure
t Check the position of the motor in the “Separate motor” display.
The positioning bars must be in the top position (1).
t Click on the [Lower] button and hold it down.
y The device is disconnected from the Palletank
®
for Mixing and the
impeller in the mixing bag. During this process the distance sensor, the
alignment pins, and the motor are lowered into the starting position and
the device gives off acoustic signals.
8.12.2 Emptying and Removing the Mixing Bag
The mixing bag in the Palletank
®
for Mixing needs to be emptied and
removed from the Palletank
®
for Mixing (see instructions for the Palletank
®
for Mixing).
8.12.3 Removing the Device from the Palletank
®
for Mixing
Requirements
The device is decoupled from the impeller and Palletank
®
for Mixing.
WARNING
Risk of injury when removing the device from the Palletank
®
for Mixing!
t Ensure that there are no people in the operating area of the device or
the Palletank
®
for Mixing.
CAUTION
Risk of crushing due to moving parts!
When removing the device from the Palletank
®
for Mixing, the device’s
guide rollers glide over the guide rails of the Palletank
®
for Mixing.
t Do not reach into the area between the guide rollers and guide rails of
the Palletank
®
for Mixing.
t Ensure that there are no people in the operating area of the device who
can reach between the guide rollers and guide rails.
CAUTION
Trip hazard from connection cable!
t While moving the device pay attention to where the connection cable is.
Lift down the motor by pressing the
arrow down button. Press the button
until the motor is positioned correctly
Separate motor
Access level: Admin
DD.MM.YYYY
HH:MM:SS
Manual mixing