Configuring the Printout
(CONFIG)
Purpose
You can configure individual printout
formats for each application.
All analysis printouts have basic factory
settings.
You can block access to the “Printout
configuration” data by assigning
a password.
Features
– Maximum items in a data record: 30
– Header, footer,
intermediate results, statistics and
information records can be configured
separately
– Printout header is output when starting
moisture analysis
– Intermediate records are output
during the moisture analysis runs by
pressing p or
at each printout interval
– Footer printed out at the end of
a moisture analysis
– Statistics are printed out while the
statistics are displayed by pressing p
(MA100 only):
– Information on phase drying is printed
by pressing p while the information
is displayed
– Printout items can be deleted
individually
– “Form feed” in a printout footer:
Select the “YDP01IS label” print mode
to configure automatic form feed up
to the beginning of the next label
Extra Functions
§ Exit “Printout configuration”:
Press the oo soft key
> Restart the application
Printout LIST or SELECT
– LIST: Output of the current printout
list
SELECT: Print currently
selectable items
§ When the select bar is on LIST or
SELECT: Press p
> Printout (example)
PRINT HEADER
LIST
====================
Blank line
GLP header
Program name
Heating param.
Standby temp.
Start parameter
End parameter
Initial weight
--------
====================
etc.
Output of All Printout Settings:
– When the select bar is on printout
overview (SETUP CONFIG.):
Press p
> Printout (example)
--------------------
13.07.1999 13:08
Mod. MA100C
Ser. no. 90706913
Ver. no. 01-38-07
ID
--------------------
SETUP
CONFIG.
--------------------
Printout header
Blank line
GLP header
Program name
Heating param.
Standby temp.
Start parameter
End parameter
Initial weight
--------
Intermediate result
Analysis time
Analysis result
Printout footer
--------
Ending time
Final weight
Final result
--------
etc.
§ Reset “Printout Configuration” to the
factory settings:
See “Printout configuration”
Set Printouts to
Factory Settings and confirm
with YES.
20