External Calibration/Adjustment with a Factory-Defined Weight
Externally calibrate weighing system and adjust using factory-defined (standard) weight
Step Key (or instruction) Display/Printout
1. Select
isoTEST function T key
2. Select “Weighing system settings”
O soft key
3. Select external calibration/adjustment
O soft key
with factory-defined weight
(tare the weighing system, if necessary)
4. Start external calibration
Start soft key
5. Place standard weight (e.g., 50 g) Load standard weight
on the weighing system
Minus sign –: weight too low
Plus sign +: weight too high
No plus/minus sign: weight o.k.
This is displayed after calibration:
(On verified analyzers, the deviation
between the readout and the conventional
measurement is displayed)
6. If you do not need to adjust
End soft key --------------------
the weighing system 07.09.1999 13:03
Mod. MA100C
Ser. no. 90805355
Ver. no. 01-38-07
ID
--------------------
External calibration
Nom. + 50.000 g
Diff. + 0.001 g
--------------------
07.09.1999 13:03
Name:
--------------------
52