Data Input Format
You can connect a computer to your weighing system to send commands via the weighing
system interface port to control functions and applications of the weighing system and
the drying unit. SBI functions for remote control are only useful, if the current device/appli-
cation mode can be read out any time.
The commands sent are control commands and may have different formats; e.g., control
commands can have up to 26 characters. Each character must be transmitted according to
the settings configured in the Setup menu for data transmission.
Format for Control Commands
Format 1: Esc ! CR LF
Format 2: Esc ! _ CR LF
Format 3: Esc ! # _ CR LF
Format 4: Esc ! # max. 20 &_ CR LF
Esc: Escape _: Underline (ASCII 95)
!: Command character CR: Carriage RETURN (optional)
#: 1 or 2 digits LF: Line FEED (optional)
&: Digit or letter max. Depends on command character, i.e.
parameter: once the max. length is
attained, the input received cut off,
rather than being rejected as with
a keyboard input.
63
Format 1
! Meaning
K Weighing mode 1
L Weighing mode 2
M Weighing mode 3
N Weighing mode 4
O Block keys
P Print
Q Beep
R Unblock keys
S Restart
Format 2
! Meaning
ars Read status
acc SBI commands only, no output
to printer
acp SBI commands with output
to printer
Format 3
!# Meaning
kF1 Soft key 1* Function depends
kF6 Soft key 6* on system state
kF7 Function key s
kF8 Function key T
kF9 Function key l
kF10 Function key k
kF11 Function key c
x1 Print weighing system model
x2 Print serial number
x3 Print weighing system
software version
x4 Print operating software version
x5 Print (GLP) weighing system
ID number
x6 Print weight set (“inventory”)
number
Format 4
!# Meaning
t Alphanumeric input
(max. character number depends
on display)
z5 Input (GLP) ID number
(20 characters max.)
z6 Input weight set number
(14 characters max.)
* numbered from right to left