Read Status: Application condition
2
)
Operating mode Action Display
1 TARE Tare the sample pan Display analysis value
2 WEIGH-IN Load the sample Display analysis value
3 WEIGH-IN Start analysis Display analysis value
4 TARE/ANALYS. BEGIN Waiting for stability Blank
5 TARE Waiting for taring Blank
to be completed
6 ANALYS. BEGIN Close the sample Display analysis value
chamber
Time delay begins
7 ANALYSIS Analysis in progress Display mode
9 ANALYS. END/CANC’LD End of analysis Display mode
10 INFO Info mode activated Menu
11 MODE Parameter mode Menu
activated
12 ID ID input activated Menu
13 STATISTICS Statistics display Menu
activated
14 PROGRAM Program memory Menu
activated
15 PAUSE Wgh.sys error ERR xxx
50, 54, 53
Data Output Format
You can output the values, displayed
in the corresponding line for moisture
analysis values, and the weight unit
with or without a data ID code.
Example: With data ID code
N + 3.4253 g
The number of characters output per
line depends on the moisture analyzer
model:
for an (optional) internal printer:
24 characters per line
YDP01ISI Label: 22 characters per line
SBI operating mode:
20 characters per line
The output format described below is
only valid for outputs of standard
measured values
(SBI: 20 characters + CR LF). If more
characters are output, 2 spaces will be
entered between the ID code I and
plus/minus sign.
Output Format with 20 Characters + CR LF
1 2 3 4 5 6 7 8 910 11121314151617181920 21 22
I I I I I I + * DD DDDDDD * UUUCR LF
***** .. ..... . ***
XX X y y z z
* ** ******
000000
I: ID code character
1
) U: Unit symbol
1
)
*: Space CR: Carriage return
D: Digit or letter LF: Line feed
1
) depends on the MA model XXX xx zz: Response to the query
Esc ars_ (“Read Status”)
2
)
Error Codes
1234567891011121314151617181920 21 22
S t a t * * * * * E R R * # # # * * * * CR LF
*: Space # # #: Error code number
ID code character I
1
)
see page 21
Read Status: Active Application XXX
2
)
MAN Moisture analysis
SET Setup menu
TST isoTEST menu
CBA Cal./adj. wgh.sys.
CHT Heater adjustment
CMS Analysis adjustment
THW Hardware test
Read Status:
Sample Chamber Status zz
2
)
O Sample chamber is
completely opened
C Sample chamber is
completely closed
OC Sample chamber is being
closed now
CO Sample chamber is being
opened now
E Undefined condition
62