Example 4: Gentle Drying with Semi-automatic Shutoff Mode (MA100 only)
The moisture content of 3.5 g of an unknown substance is to be determined. The sample to be analyzed is known to contain surface moisture
and the moisture of two bound fractions. The three moisture contents are to be individually quantified by drying.
Settings (different from the factory settings):
Program number: 4
Program name: substance N
Heating program: phase drying
Termperature 1: 50°C
Time 1: 4 minutes
Temperature 2: 105°C
Time 2: 6 minutes
Final temperature: 200 °C
End of analysis: Time: 4 minutes
Printout footer: Phase results (for input sequence see the chapter entitled “Configuring the Moisture Analyzer”)
Step Key (or instruction) Display/Printout
1. Turn on the moisture analyzer e key Sartorius logo is displayed
Self-test runs
The above mentioned parameters Enter the parameters:
for drying program 4 see Example 1
“Substance N”
have already been set
2. Select program display
Prog. soft key
3. Select program “4” q soft key
4. Load the program “Substance N”
Load soft key
5. Confirm load prompt
Yes soft key
45
Info TareID Stat. Mode Prog.
P3 100 C 4mg/30s
TARE: Tare sample pan
+ 0035g
Info TareID Stat. Mode Prog.
P4 50/105/200 C Time
TARE: Tare sample pan
+ 0004g

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius MA100 and is the answer not in the manual?

Sartorius MA100 Specifications

General IconGeneral
Capacity100 g
Readability0.1 mg
Weighing pan size90 mm diameter
DisplayLCD
InterfaceRS232
Measurement PrincipleElectromagnetic force compensation
Repeatability (standard deviation)0.0001 g (0.1 mg)
Weighing Range0 to 100 g
Power Supply50/60 Hz

Related product manuals