Digital Input/Output Ports
Monitoring, Remote Control
You can monitor and remote-control
the moisture analyzer. For this purpose,
you should know the operating state
of the unit.
The following external devices can be
used here:
Status display with digital input ports
Process logic controllers
Computer with communications port
Via the communications port, the status
of the moisture analyzer and therefore
the operating state of analysis can
be read out. The moisture analyzer is
remote-controlled by the functions
used to activate the key functions.
65
Functions for Remote Control
(see also “Data Input Format” and
“Data Output Format”):
Read out the moisture analyzer status:
ESCars_
Block/unblock the keys on the moisture
analyzer:
ESCO / ESCR or on the moisture analyzer
via Setup: Device
Activate key functions:
ESCP, ESCkF1_ to ESCkF11_
The respective operating state of the mois-
ture analyzer will be transmitted to the
digital output ports according to the chart
“Operating State of Analysis” (see below).
The moisture limits for the control
function at the ANALYSIS END (in toler-
ance, < min., > max.) are to be entered
under soft key
Mode: Limits for
control function.
Remote control via the universal remote
control switch input (pin 15):
Set the universal remote control switch
function to “Function key F1” (“Setup:
Device: Extra functions: Ext. switch”)
$ Block/unblock the keys on the moisture
analyzer, if necessary (“Setup: Device:
Keys: Block”)
$ For complete control and analysis using
the F1function key, set the analysis start
mode to “Fully automatic, with (with-
out) stability” (Prog.xx: Start analysis)
If a bar code scanner/additional keypad
is set, remote control is no longer
possible via the universal remote control
switch input.
Operating State of Analysis
State Pin 16 Pin 17 Pin 18 Pin 19 Description
No analysis in progress 0 0 0 0 No drying program (Mode e.g., Off,
Standby, Setup, isoTEST, Prog., Err xxx)
TARE/ WEIGH-IN 0 0 0 1 Drying program before analysis runs, but no
Menu display values displayed: Mode, Info, ID, Statistics
TARE pause 0 0 1 0 Waiting for Tare soft key
TARE Waiting for taring to be
completed 0 0 1 1 Tare soft key pressed; waiting for stability
WEIGH-IN Place the sample 0 1 0 0
WEIGH-IN start analysis 0 1 0 1
ANALYSIS started 0 1 1 1 e.g., Start analysis, Mode, Info
ANALYSIS display 0 1 1 0
ANALYSIS END (within tolerance) 1 0 0 0/1 Analysis value display/Mode, Info, Statistics
ANALYSIS END (< min.) 1 0 1 0/1 Analysis value display/Mode, Info, Statistics
ANALYSIS END (> max.) 1 1 0 0/1 Analysis value display/Mode, Info, Statistics
ANALYSIS CANCEL 1 1 1 0/1 Analysis value display/Mode, Info, Statistics