Specifications for Models with EC type-approval certificate
Model MA100C-0CE230V1, MA100H-0CE230V1
Dryer functions:
Heating element Ceramic IR heater (MA100C) or round halogen lamp (MA100H)
Temperature range 30 – 230°C
86 – 392°F
Temperature increments adjustable in 1°C increments
Temperature adjustment with YTM03MA temperature adjustment set
Weighing Functions:
Accuray class
1)
k
Max. weighing capacity
1)
100 g
Scale interval d
1)
1mg, 0.01%/0.001 % moisture content
Tare range (subtractive) g<100 % from max. weighing capacity
Verification scale interval e
1)
0.001 g
Min. capacity
1)
0.1g
Ambient temperature range +15 °C ... +25 °C
Sample pan dimensions Ø 90 mm
1)
according to Council Directive 90/384/EEC on non-automatic weighing instruments used within the European Economic Area
Drying parameters:
Drying programs standard, quick,
gentle, 3 temperature phases
Drying time 6 sec. to 999 min
Number of programs 30
Shutoff criteria Fully automatic, semi-autom.,
asap, time (3x 999 min.),
manual
Display for analysis results Moisture (calculated value),
dry weight, RATIO,
weight loss,
residual weight (g or g/kg)
Analyzer (hardware):
Dimensions (L x W x H) 350 x 453 x 156 mm
Net weight, approx. 8 kg
Voltage 230 V or 115 V selectable by replacing
the heating unit, –15% ... + 10%
Frequency 48 – 60 Hz
Fuses 2 (neutral conductor/phase), 6.3 AT, 5 x 20 mm
Operating temperature range +10 ... +30°C (50°F – 86°F)
Power consumption 700 VA max.
Built-in interface RS-232C
Format: 7 or 8 bit ASCII, 1 start bit, 1 or 2 stop bits
Parity: Space, odd or even
Transmission rates: 150 to 19,200 baud
Handshake: Software or hardware
Digital input: 1, adjustable function
Digital outputs: 4, operating state of analysis
73