Service
Regular servicing by a Sartorius
technician will extend the service life of
your moisture analyzer and ensure its
continued weighing accuracy. Sartorius
can offer you service contracts, with
your choice of regular maintenance
intervals ranging from 1 month to 2
years.
The frequency of the maintenance
intervals depends on operating
conditions and user tolerance
requirements.
Repairs
Repair work must be performed by
trained service technicians. Any
attempt by untrained persons to
perform repairs may lead to hazards for
the user.
Cleaning
! Make sure that no dust or liquid enters
the moisture analyzer housing
! Do not use any aggressive cleaning
agents (solvents, abrasive cleaning
agents, etc.); clean the moisture ana-
lyzer using a piece of cloth which has
been wet with a mild detergent (soap)
only
§ Unplug the power cord from the wall
outlet (mains supply)
If you have a cable connected to the
interface port, unplug it from the
moisture analyzer
$ The base plate of the sample chamber,
the shield disk and the pan support can
be removed for cleaning
§ Carefully remove any sample residue
/spilled powder by using a brush or a
hand-held vacuum cleaner
§ After cleaning, wipe down the analyzer
with a soft, dry cloth
Cleaning the Heating Unit and
Temperature Sensor
!Danger: The terminals of the
heating unit are under live current
§ To disconnect from AC power, unplug
the power cord from the wall outlet
(mains)
If you have a cable connected
to the interface port, disconnect
it from the moisture analyzer
§ Slide out the heating unit to remove
§ Carefully remove any residue from the
temperature sensor
§ Place the disassembled heating unit on
an even surface and use a commercially
available cleaning agent for glassware
to clean the ceramic heating element or
the protective glass of the halogen lamp
Replacing the Fan Air Filter
$ Remove any dust from the fan air inlet-
located on the bottom of the moisture
analyzer
$ Check the air filter periodically and
replace it if necessary (see Accessories)
§ Place the moisture analyzer on its left
side
§ Unscrew the cover plate
§ Pull out the air filter holder
§ Open the air filter holder and remove
the air filter
70
Care and Maintenance