Internal Calibration/Adjustment (MA100 only)
Inside the moisture analyzer is a built-in motorized calibration weight.
Step Key (or instruction) Display/Printout
1. Select
isoTEST function T key
2. Select “Weighing system settings”
O soft key
function
3. Select “Internal calibration/ q soft key twice
adjustment”
O soft key
4. Start internal calibration sequence
Start soft key
Internal calibration weight is
applied automatically
The weighing system is calibrated
The internal calibration weight is
removed from the weighing system
This is displayed after calibration:
6. If you do not need to adjust
End soft key Calibration printout is generated
the weighing system
Otherwise, adjust the weighing system
Start soft key
The weighing system is adjusted
This is displayed after adjustment:
Adjustment printout is generated
54