Step Key (or instruction) Display/Printout
7. Open the sample chamber k key
Position a new sample pan
8. Tare the sample pan
Tare soft key
9. Distribute approx. 2.5 g of
fine ceramic material uniformly
on the sample pan
Close the sample chamber k key
10. Start the drying program
Start soft key
After 2 seconds’ delay, --------------------
the header for the moisture 14.07.1999 14:10
analysis will be printed Mod. MA100C
Ser. no. 90805355
Ver. no. 01-38-07
ID
--------------------
Prg 2 FINE CERAM MAT
Heating QUICK
Fin. temp. 200 ’C
Stdby temp. OFF
Start W/STABIL.
End AUTOMATIC
IniWt + 2.584 g
--------------------
The current moisture loss
will be displayed afterwards
Fully automatic shutoff of the
drying process, if no further
weight loss is registered
(in this case: after 9 minutes)
The footer for the moisture --------------------
analysis is printed Analys.Time 9.0 min
FinWt + 2.246 g
Res + 86.89 %R
asap
Difference 1.2 %
Interval 24 sec
--------------------
42
Start TaraID Stat. Mode Prog.
P2 200 C Auto.
WEIGH-IN: Load sample pan
+ 0000g
Start ID Stat. Mode Prog.
P2 200 C Auto.
WEIGH-IN: Start analysis
+ 2584g
Start
Start ID Stat. Para. Prog.
P2 200 C Auto.
ANALYS.BEGIN: Delay start 2 sec
+ 2584g
Abbr.
S
Start ID Stat. Mode Prog.
P2 200 C Auto.
ANALYSIS: 37 C 0.8min ----mg/24s
+ 9815%R
Cancel
S
Start ID Stat. Mode Prog.
P2 200 C Auto.
ANALYSIS END: 198 C 9.0min
+ 8689%R
Next
S