Step Key (or instruction) Display/Printout
7. Open the sample chamber k key
Position a new sample pan
8. Tare the sample pan
Tare soft key
9. Distribute approx. 2.5 g of
fine ceramic material uniformly
on the sample pan
Close the sample chamber k key
10. Start the drying program
Start soft key
After 2 seconds’ delay, --------------------
the header for the moisture 14.07.1999 14:10
analysis will be printed Mod. MA100C
Ser. no. 90805355
Ver. no. 01-38-07
ID
--------------------
Prg 2 FINE CERAM MAT
Heating QUICK
Fin. temp. 200 ’C
Stdby temp. OFF
Start W/STABIL.
End AUTOMATIC
IniWt + 2.584 g
--------------------
The current moisture loss
will be displayed afterwards
Fully automatic shutoff of the
drying process, if no further
weight loss is registered
(in this case: after 9 minutes)
The footer for the moisture --------------------
analysis is printed Analys.Time 9.0 min
FinWt + 2.246 g
Res + 86.89 %R
asap
Difference 1.2 %
Interval 24 sec
--------------------
42