Step Key (or instruction) Display/Printout
The sample is heated to
140°C and dried for at least
5 minutes (phase 1).
If a constant weight has not been
reached after 5 minutes, the
drying process is continued
until a constant weight is reached
(in this case: after 9.5 minutes).
After the constant weight has been Analys.time 9.5 min
reached, the footer for the FinWt + 0.836 g
moisture analyzer is printed. --------------------
Res1 + 25.74 %l
Res2 + 0.00 %l
Res3 + 18.57 %l
--------------------
C-Res + 44.31 %l
--------------------
50