Picus
®
2 Operating Instructions 53
Technical Data
14.2 Ambient Conditions
Unit Value
Laboratory, for indoor use only
Pollution level according to IEC 61010-1 II
Maximal height above sea level m 2000
Suitable for protection class
Protection class of the device, as per DIN EN 60529-1 IP 40
Temperature
In operation °C +15 – +30
During performance review °C +15 – +30
In storage | during transport °C -20 – +40
Relative humidity, in operation
At temperatures up to 31°C, maximum % 80
No potentially explosive areas
Other properties
Store dry
14.3 Electrical Data
14.3.1 Power Supply
Unit Value
Alternating voltage V
AC
100 – 240 (± 10 %)
Frequency Hz 50
Current consumption, maximum A 0.125
Power supply only permitted through mains connection cable provid-
ed by Sartorius
Protection class, as per IEC60950-1,
(reinforced insulation between mains and secondary circuit)
II
Surge category according to IEC 60664-1 II
Connection cable
Connection cable with micro-USB port
Length m 1,8