Picus
®
2 Operating Instructions 59
Technical Data
14.12 Performance Specification
14.12.1 Models LH-747021 | LH-747041 | LH-747061 | LH-747081 | LH-747101 | LH-747111
Channel
Colour of the
operating button
Volume range
Increment
Test volume
Mode*
Systematic error ±
Random error ± ±**
Unit Unit Unit Unit Unit Unit
µL µL µL µL % µL % µL
Value Value Value Value Value Value Value Value
LH-747021 1
n
0.2–10 0.01 10 P 1.0 0.100 0.4 0.040
5 P 1.2 0.060 0.7 0.035
1 P 3.0 0.030 2.0 0.020
1 D 6.0 0.060 7.0 0.070
LH-747041 1
n
5–120 0.10 120 P 0.5 0.60 0.15 0.18
60 P 0.7 0.42 0.2 0.12
12 P 2.0 0.24 1.0 0.12
5 P 5.5 0.275 2.5 0.125
12 D 4.0 0.48 4.0 0.48
LH-747061 1
n
10–300 0.20 300 P 0.5 1.50 0.15 0.45
150 P 0.6 0.90 0.2 0.30
30 P 1.5 0.45 0.8 0.24
10 P 5.0 0.50 2.4 0.24
30 D 3.0 0.90 3.0 0.90
LH-747081 1
n
50–1,000 1.00 1,000 P 0.45 4.5 0.15 1.5
500 P 0.6 3.0 0.2 1.0
100 P 2.0 2.0 0.5 0.5
50 P 4.0 2.0 1.0 0.5
100 D 2.5 2.5 2.0 2.0
*P = Pipetting mode
D = D mode for Multi Dispensing.
The systematic and random measurement deviations listed here apply for ten measurements taken at 10% of the
nominal volume.
** Note: the values listed for the systematic and random measurement deviations can only be achieved under
strictly controlled conditions during type examinations in accordance with ISO 8655. Due to continuous product
development, Sartorius reserves the right to change these values without prior notification.