3 Initial Preparation
3.2 Connection to a Host System
Page 26
BP-200 plus
Art. no.: 2.510701 rev.: c
3.2.4 AT-10 plus Connections
3.2.5 AT-10 plus Settings
Defining the trigger pulse
In the system menu select:
Menu Key > Peripherals > QRS Trigger
The signal is a TTL logic level of 5 V. For use with the BP-200 plus make the following
settings:
• QRS Trigger: On
• Amplitude: +5 V
• Baseline: 0 V
• Duration: 50 ms
• Delay: 10 ms
Defining the BP Unit
In the settings menu select:
• Menu Key > Peripherals > Blood Pressure Unit
A number of options are available (including manual and off). Select BP-200 plus. If
BP-200 plus is not shown select BP-200.
AT-10 plus BP-200 plus
RS-232 Cable Ass.
Art. No. 4.520373
Trigger Cable Ass.
Art. No. 4.520564
RS-232
ERGO
(side
panel)
RS-232
(back
panel)
Comm. Network SCM Modem Peripherals
Identification System Config. Software
System Menu
RS-232
QRS
Duration (ms)
Delay (ms)
Amplitude (V)
Baseline (V)
The BP is taken one minute before the end of every exercise stage defined in the
protocol (see the AT-10 plus user guide).