6 Maintenance
6.1 Maintenance Intervals
Page 42
BP-200 plus
Art. no.: 2.510701 rev.: c
6 Maintenance
6.1 Maintenance Intervals
All maintenance tests and intervals are subject to verification.
All software controlled devices have undergone a software risk analysis to minimise
any hazards connected to software defects.
Maintenance work not described in this chapter may only be accomplished by a
qualified technician authorised by SCHILLER.
The following table indicates the intervals and responsibilities of the maintenance
work required.
Interval Maintenance Responsible
Every 6 months Visual inspection of the unit, cables. tubes and cuff
Keys test (see next page).
« User
Every 12 months All checks performed at the six month interval.
Functional and safety checks and inspections according to the in-
structions in the service handbook.
NIBP calibration check.
ECG check.
•SpO
2
check.
« Service staff authorised by
SCHILLER