6 Maintenance
6.7 BP-200 plus Error Messages
Page 48
BP-200 plus
Art. no.: 2.510701 rev.: c
6.7 BP-200 plus Error Messages
The following is a list of the error message that can appear on the BP-200 plus.
6.7.1 Error Message Table
No. Message Cause Remedy
51 Battery low
Backup batteries voltage too
low to take a measurement.
« Replace with fully charged accumulators.
52 Valve/pump
Valve or pump is defective; air
leak.
« Check pneumatic system. Check that tube is se-
cured and that the connectors at both ends are se-
curely in place. Check for holes or air leaks in the
cuff bladder and pressure hose; contact service de-
partment.
120 Signal Disturbed
Disturbances in measurement
signal.
« Perform measurement in quiet environment; avoid
moving the arm during a measurement.
122 Cuff pressure
Too much residual air in cuff
bladder.
« Ensure that cuff is completely decompressed before
commencing a measurement.
123 No cuff
No pressure (<3 mmHg) 10 sec
after inflation; leak in the cuff or
hose.
« Connect cuff; check hose connections.
124 Measur. break
Measurement stopped manual-
ly.
126 Measurement time
Measurement time too long;
hose, cuff or pump is defective.
« Check hose connections. Check hose and bladder
for defects. Contact service department.
127 Cuff loose
Pressure too low (<10 mmHg)
10 sec after inflation.
« Check cuff; if necessary tighten it more firmly.
128 Pump time
Pump-up time too long (50/60
sec, depending on patient type,
C/A); hose, cuff or pump is de-
fective.
« Check cuff and hose. Replace accumulators; con-
tact technical service.
130 Overpressure
Device automatically shut off by
relief valve.
« Maximum pressure reached.
140 Meas. invalid Evaluation impossible (KOR).
« Check position of microphone; repeat measurement.
150 Meas. invalid Evaluation impossible (OSC).
« Check cuff; avoid moving the arm.
180 No signal
Pressure reached 70 mmHg
and no Korotkoff signal detect-
ed.
« Connect microphone. Check position of microphone.
181 Weak signal
Pressure reached 50 mmHg
and no Korotkoff signal detect-
ed.
« Place microphone over brachial artery; ensure that
the microphone is correctly orientated (dark side to-
wards upper arm).
182 Signal disturbed
Too much interference in Ko-
rotkoff signal.
« Perform measurement in quiet environment; avoid
moving the arm during measurement.
190 No signal
Pressure reached 50 mmHg
and no pulse detected (OSC).
« Check cuff.
191 Weak signal
Less than 8 pulse beats detect-
ed (OSC).
« Check cuff.
192 Signal disturbed
More than 200 pulse beats de-
tected (OSC).
« Avoid moving the arm during measurement.
198 DIA not detected
Minimum pressure reached
without diastolic value detected
(OSC).
« Avoid moving the arm during measurement.