Page 49
Art. no.: 2.510701 rev.: c
Maintenance 6
User Guide BP-200 plus Error Messages 6.7
BP-200 plus
199 SYS not detected
• Minimum pressure reached
without systolic value detected
(OSC).
« Avoid moving the arm during measurement.
200 No pulse • No pulse detected (OSC).
« Repeat measurement; check cuff.
201 Pulse > max.
• Too many pulse beats detected
(OSC).
« Perform measurement in quiet environment; avoid
moving the arm during measurement.
202 Pulse < min.
• Too few pulse beats detected
(OSC).
« Repeat measurement; check cuff.
210 No pulse
• No pulse detected (KOR).
« Connect microphone; Check position of microphone.
Ensure that the microphone is correctly orientated
(dark side towards upper arm).
211 Pulse > max.
• Too many pulse beats detected
(KOR).
« Perform measurement in quiet environment; avoid
moving the arm during measurement.
212 Pulse < min.
• Too few pulse beats detected
(KOR).
« Place microphone over brachial artery; ensure that
the microphone is correctly orientated (dark side to-
wards upper arm).
238 Min. pressure
• Minimum pressure reached
without measurement result.
« Repeat measurement.
239 Max. pressure
• Maximum pressure reached
without measurement result.
« Repeat measurement.
255 Unit error • Device is defective.
« Contact technical service.
No. Message Cause Remedy