6 Maintenance
6.8 Trouble Shooting
Page 50
BP-200 plus
Art. no.: 2.510701 rev.: c
6.8 Trouble Shooting
Trouble shooting procedures are subject to verification.
Error Cause Remedy
NIBP
No cuff connected or not correctly
fitted.
Incorrect Cuff size
Inaccurate Measurement
Unit always takes two measure-
ments
Pump is not running
Pressure not reached
« Check the cuff position.
« Change cuff.
« Check cuff position.
« Check microphone position.
« Check microphone cable.
« Check for leaks in system.
« Check initial pressure.
« Check maximum initial pressure.
« Pump is mechanically blocked, electrical failure.
« Leak in system - check all hoses and connections.
« Check NIBP settings.
« Check cuff and connection.
SpO
2
Low perfusion or other
problems
Bad sensor positioning
•SpO
2
sensor failed or disconnect-
ed
« Check the sensor and reapply.
« Check the contact between the sensor and the patient.
« No sensor connected
« Damaged sensor connected.
« Interferences detected.
« Sensor not attached to patient.
« Too much ambient light for reliable measurement.
« Unknown sensor connected.
« Hardware (board) failure.
« Replace the sensor.
ECG
ECG cable disconnected
Electrode lose/defective
Incorrect settings
« Connect the ECG cable.
« Check and reapply/replace electrodes.
« Check ECG settings.