2.3.1 Overview on the configurable settings
SCHILLER's service centre can configure the following parameters:
Important!
The “FRED CO” software is only available to service centres authorised by Schil-
ler
.
Modifications that can be made via "FRED CO" are only performed if requested
by the customer, or if required by legal requirements.
These modifications need to be registered in the device documentation as well
as communicated to all users.
Configurable parameters (by means of FREDCO®)
Volume of language output
Energy level for 1st, 2nd and 3rd shock (separate settings for adults and children)
Manual or automatic start of the ECG signal analysis (only in semi-automatic op-
eration)
Activation/deactivation of the 16.7 Hz filter
a
Sound recording Yes/No
a. Sound recording — The owner needs to inform users that the device
records ambient noise during interventions.
Number of chest compressions for children (15 or 30)
Self-test frequency (daily or weekly)
Entering the device name
Choice between "continuous chest compressions" or "alterning chest compres-
sions/breathes" during CPR cycles
Time and date On
Update the software/change the device language