EasyManua.ls Logo

Schiller FRED easy - Trouble Shooting; Error Messages

Schiller FRED easy
94 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 73
Maintenance 7
User Guide Trouble Shooting 7.5
Art. no.: 0-48-0013 Rev. m
FRED easy
7.5 Trouble Shooting
7.5.1 Error messages
If it is not possible to get the device back into operating condition within a reason-
able period of time, continue cardiopulmonary resuscitation until the rescue service
arrives.
If the device switches to Limited mode (CPR) due to an error, the alarm messages
may differ.
Forced shutdown procedure
If the device cannot be switched off via normal OFF procedure (press and hold the
green button for approx. 3 seconds), remove the battery and inserting again.
Error message Possible cause Remedy
xxx =
–ADC
EEPROM
–RTC
–LCD
–OKI
–DSP
SHOCK BUTTON
CHARGE
DEFIBRILLATOR
COMMUNICATION
DEFIBRILLATOR
5 WEEKS
CHARGE 150J
Technical errors that may occur
during function tests:
Switch the unit off and then on again to con-
firm. If the message still appears, the device
must be repaired immediately.
xxx =
–CPU DEFI
CRC DEFI
SAFETY DEFI
REF VOLTAGE DEFI
ADC DEFI
DEFI DISCHARGE
DEFI EPROM
DEFI SHOCK KEY
Technical problems that can oc-
cur during defibrillation
Switch the unit off and then on again to con-
firm. If the message still appears, the device
must be repaired immediately. Continue
cardiopulmonary resuscitation until the res-
cue service arrives.
Error
INFO: XXX
PERFORM
CHEST COMPRESSIONS
TEST FAILURE
XXX
Error
INFO: XXX

Table of Contents

Related product manuals