5Versions
5.3 Online version
Page 54
Art. no.: 0-48-0013 Rev. m
FRED easy
5.3 Online version
The internal memory is able to record:
2 hours of ECG signal,
2 hours of ambient noise (if the sound recording parameter is activated using
FREDCO®),
500 events concerning the intervention.
Once the internal memory is full, the symbol starts blinking and no more data can
be stored. Stored data will be cleared once it has been transmitted.
Operating principle
This version is intended to ensure communication between the device and the LifeD-
ataNet® web server.
Just as with the Ethernet version, this version is able to send intervention data (stored
on the internal memory) as well as detailed device configuration data to the LifeDa-
taNet® web server. It also allows a total remote control of the FRED easy® using Life-
DataNet® (see table below).
Communication takes place via a network infrastructure using the SNMP (Simple Net-
work Management Protocol) communication protocol.
Fig. 5.9 Online version with docking sta-
tion
The SD card is not removable in the FRED easy® Online version.
All events transmitted by the device are listed on page 48.
Overview of actions that may be performed remotely, using LifeDataNet®:
Monitoring of battery capacity and self-test result
Update of device’s software
Modification of device configuration
Modification of device language
Triggering of audible alarms or visual indications on the devices
Viewing the device status
Geographic location of devices (Google Maps)