7 Maintenance
7.1 Maintenance Intervals
Page 70
Art. no.: 0-48-0013 Rev. m
FRED easy
7.1.2 Green indicator
7.1.3 Functional check
Points to inspect: Visually inspect the device and the accessories (see 7.1.1 Visual inspection of the
device and accessories).
Check for proper functioning.
Measure the leakage current.
Measure the energy delivered at 50 Ohms.
7.1.4 Internal backup battery
The internal backup battery must be replaced at least every 6 years by SCHILLER
technical support department or an authorised distributor.
Fig. 7.1 Blinking green indicator
If the device is defective or if problems have been detected by the device during the
self-test (green indicator not blinking), the device must be repaired before use.
Patient hazard — If the device's behaviour differs from the description given in this
user guide, the device is defective and must be repaired.
In case of intensive use of the device, SCHILLER recommends that these in-
spections be performed at shorter interval.
The regulations in force in each country regarding inspection frequency must be
observed (if shorter intervals than those recommended by SCHILLER are im-
posed).
The old battery must be recycled in accordance with local regulations.