6Options
6.1 ECG display
Page 62
Art. no.: 0-48-0013 Rev. m
FRED easy
6Options
6.1 ECG display
When the "ECG Display" option is activated, the ECG signal obtained by the elec-
trodes is displayed on the screen (see Fig. 6.1 ECG display).
Fig. 6.1ECG display
6.2 Metronome
When the "Metronome" is enabled, the device sets the rhythm for chest compres-
sions during CPR at 100 bpm.
The FREDCO® software program allows you to:
set the metronome frequency to a value between 80 and 150 bpm (in steps of
5 bpm).
the metronome can be set to signal to the user to give 2 rescue breaths (the
metronome is stopped for 8 seconds after 30 chest compressions).
PRESS GREEN BUTTON
The "ECG Display" option is only available for the FRED easy® semi-automatic.
This option is automatically activated when the FRED easy® is switched to manual
mode.
When the ECG signal is displayed, messages are only displayed on one single
line.