5Versions
5.1 SD card version
Page 48
Art. no.: 0-48-0013 Rev. m
FRED easy
5 Versions
5.1 SD card version
Operating principle
This version records all defibrillation-related events on a removable SD card.
The SD card is able to record:
2 hours of ECG signal,
2 hours of sound recording (if this parameter has been activated using FRED-
CO® and unless ambient noises are too loud).
500 events concerning the intervention (see table below).
Equipment damage hazard —
The memory card slot must always be covered with the rubber cover. This is to
prevent moisture penetrating the device.
Always turn off the device before inserting or removing the SD card.
Do not insert the Ethernet adapter into the SD card slot.
Malfunction hazard
The SD card must only be used in one single device.
Before being used in another device, the SD card must be reformated with the
Schiller data viewing software; otherwise, the recorded information will be incor-
rect.
Only use SD cards supplied by SCHILLER.
Overview of recorded events, with date and time:
Power on
Movement detection (beginning and end)
Operating mode
Start of analysis
Analysis result
Defibrillator charging
Defibrillation shock
Defibrillation shock cancelled
Internal discharge
a
a. This event is only recorded when the defibrillator has switched to "limited mode"
due to a technical problem.
Electrode alarm
"Battery low" alarm
Critical error
Power off