4 Defibrillation
4.6 Manual Defibrillation (option)
Page 46
Art. no.: 0-48-0013 Rev. m
FRED easy
Step 3 Shock delivery
6. Deliver the shock by pressing the button .
After the shock delivery, proceed with Step 2 Charging the energy.
Finishing the
therapy
See 4.8 Finishing the therapy.
4.6.3 Switching to semi-automatic operational mode
To return to semi-automatic mode, turn off the FRED easy® and leave it turned off for
at least 5 minutes before switching it on again.
As soon as the energy is charged, the device prompts the user to deliver the shock by
pressing the orange button . The orange button remains lit until the shock is
delivered.
Shock hazard!
Only deliver a shock when a shockable heart rhythm is present.
Do not, under any circumstances, touch the patient during shock delivery.
Make sure that the patient does not touch any conducting objects.
Fig. 4.12 Button to deliver the shock