3 Initial operation
3.2 Battery monitoring
Page 26
Art. no.: 0-48-0013 Rev. m
FRED easy
3.2.3 Battery depleted during use, limited mode (CPR)
If a depleted battery is detected while the device is in use, a message is displayed
prompting the user to replace the battery and to perform CPR. An audible signal is
emitted and the indicator remains off until the battery is replaced (or charged).
The message stays displayed until the battery is replaced.
Patient hazard — Defibrillation is no longer possible if a depleted battery is detected.
The battery needs to be replaced immediately.
REPLACE BATTERY
30 CHEST COMPRESSIONS THEN
2 RESCUE BREATHS
Depleted battery during self-test
When a depleted battery has been detected, the green indicator is off and an au-
dible signal is emitted.
When the device is next switched on, a message is displayed, prompting the user
to replace the battery and continue with CPR until the device is fully operational
again.
Depleted battery after the battery has been inserted
If a battery inserted in the device is identified as depleted, an audible signal is emit-
ted, the green indicator is turned off and a message indicating that the battery must
be replaced is displayed.