Page 25
Initial operation 3
User Guide Battery monitoring 3.2
Art. no.: 0-48-0013 Rev. m
FRED easy
3.2.2 Low battery capacity during use
Despite the acoustic and written warnings, the device can still be used as normal
and is still able to perform defibrillations.
Always switch off the device before removing the battery.
The remaining battery capacity depends on the use and ambient conditions.
Fig. 3.3 The green indicator is not lit
If the battery capacity falls below the "low" threshold while the device is in use, the
green indicator is turned off, the symbol starts blinking on the display and the
device emits an audible signal.
These warnings are issued until the battery is replaced (or recharged). The battery
must be replaced as quickly as possible.
Fig. 3.4 Low battery capacity
Low battery capacity during self-test or after the battery has been inserted
If low battery capacity is detected during a self-test or when the battery is inserted,
the device emits an audible signal and the green indicator is not lit until the battery
is replaced (see Fig. 3.3 The green indicator is not lit).
DO NOT TOUCH THE PATIENT
ANALYSING!