8 Technical Data
8.3 Defibrillation pulse
Page 84
Art. no.: 0-48-0013 Rev. m
FRED easy
Shock delivery
With the orange button (in semi-automatic or manual mode)
Via disposable pads applied to the patient in an anterior-anterolateral or anteri-
or-posterior position
Safety discharge when:
A non shockable rhythm has been detected
The shock is not delivered within the 20 seconds after charging
An electrode problem is detected
Battery voltage is insufficient
The device is defective
The device is turned off.
Defibrillation pad connection
BF type
Defibrillation electrodes
Adult pads
Paediatric pad
Electrode cable, 2 m in length
80 cm² active surface
42 cm² active surface
VF/VT detection
Conditions for ECG analysis
Shock recommendation
VF/VT detection is based only on the ECG signal.
Minimum amplitude for signals suitable for analysis > 0.15 mV, signals < 0.15 mV
are considered asystole.
For VF and VT (VT > 150 B/min).
VF VT Non shockable
rhythms
NSR/N/Asystole
Shock 148 100 2
No Shock 2 2 1395
Performance criteria
Sensitivity > 90 % Sensitivity > 75 % Specificity > 95 %
Observed performance
98.67 % 98.04 % 99.86 %