Instructions for Use
Resting & Stress ECG with seca
ct330/331 and seca diagnostic 5.9
seca diagnostic
The device must only be used with the supplied seca software (seca
diagnostic).
As the operator you are responsible for ensuring regular data backups
(patient databases, evaluations etc.) and system backups. We
recommend that you backup the data at the latest before new
installations, updates and far-reaching system configurations.
seca diagnostic new installations, updates and system configurations
may only be performed by your authorised seca dealer.
Only change data generated in seca diagnostic within seca diagnostic
itself and not in your surgery IT system or your hospital information
system (HIS). seca does not accept any responsibility for any changes
to data in your IT system or your HIS which were made after the export
from seca diagnostic.
To ensure the safe operation of seca diagnostic, deactivate the
screensaver and energy management options on your PC. Set up your
operating system in such a way to prevent the PC from being switched
off either accidentally or automatically during the examination (standby
mode/idle mode).
Allocation of case and job numbers
If case or job numbers are manually entered into the system or they are
changed in the system, there is a risk of confusing patients and
subsequent misdiagnosis if an incorrect entry is made by a user. Always
make sure that case or job numbers are entered correctly!
Scanning or manually entering patient, case or job numbers does not
relieve the user of the obligation to check the patient to be physically
treated.
Data management in seca diagnostic: Assign evaluation
If an examination was conducted with incorrect patient data, the
evaluation can be subsequently allocated to the correct patient. Make
sure that the evaluation is definitely allocated to the correct patient.
Incorrect allocation can lead to misdiagnosis. Please note that data
which has already been exported to an external system (e.g., surgery IT
system) cannot be changed.
seca diagnostic is preset with the Assign evaluation function
deactivated; however, it can be reactivated via user rights if necessary.
Only the Supervisor can configure user rights. If the Assign evaluation
function is activated, it can be found in the evaluation search or in open
evaluations in the Options menu.
We recommend configuring user rights in seca diagnostic so that only
authorised persons can execute the Assign evaluation function.